產品列表PRODUCTS LIST
乙酸戊酯CH3CO2(CH2)4CH3氣體檢測管210U PENTYL ACETATE 10-200ppm,主要用途:用作溶劑、稀釋劑,制造香精、化妝品、人造革、膠卷、火藥等。健康危害: 對眼及上呼吸道粘膜有刺激作用,可引起 結膜炎、鼻炎、咽喉炎等,重者伴有頭痛、嗜睡、胸悶、心悸、食欲不振、惡心、嘔吐等癥狀。皮膚長期接觸可致、皮炎或濕疹。有的可發(fā)生貧血和嗜酸性粒細胞增多。燃爆危險:本品易燃,具刺激性。
更新時間:2024-12-26
在線留言乙酸戊酯CH3CO2(CH2)4CH3氣體檢測管210U PENTYL ACETATE 10-200ppm
Tube No.210U
1. PERFORMANCE
1) Measuring range :10-200 ppm
Number of pump strokes 3(300ml)
2) Sampling time :4.5 minutes/3 pump strokes
3) Detectable limit :4 ppm
4) Shelf life :2 years
5) Operating temperature :10~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 3 pump strokes
8) Colour change :Pale yellow→Pale blue(more than 20 ppm)
〃 →Dark brown(20 ppm or less)
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
Chromium oxide is reduced.
CH3CO2CH2(CH2)3CH3+Cr6++H2SO4→Cr3+
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
10
30
50
100
150
200
ppm
300ml
GAS To pump
DEHUMIDIFIER(White)
210U PENTYL ACETATE
Substance Interference Coexistence
Alcohols Similar or Blue stain is produced. Higher readings are given.
Esters 〃〃
Ketones 〃〃
Aliphatic hydrocarbons 〃〃
Aromatic hydrocarbons The accuracy of readings is not affected. Whole reagent is changed to Brown.
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 10℃ 20℃ 30℃ 40℃
(ppm) (50°F) (68°F) (86°F) (104°F)
200 240 200 160 120
150 180 150 130 100
100 120 100 80 70
50 60 50 40 30
30 36 30 24 18
10 10 10 8 6
上一篇:113SB戊烷CH3(CH2)3CH3氣體檢測管113SB PENTANE 50-1000ppm
下一篇:105SD戊胺CH3(CH2)3CH2NH2氣體檢測管105SD PENTYL AMINE 2-22ppm
返回列表>>相關產品